What the pad looks for
Ascorbic acid is vitamin C. This pad indicates its concentration band in the urine sample. It is not a test of your body’s vitamin C stores or nutritional status.
Why vitamin C matters to testing
Published studies show that urinary vitamin C can interfere with some urine dipstick reactions and contribute to false-negative readings. The affected reactions and the amount of interference depend on the dipstick chemistry. The 14-in-1 chart notes that you want this pad to read zero for correct readings on the other tests.
What it cannot tell you
A vitamin C reading does not prove that another pad is affected or tell you how to correct its value. Research on other dipstick brands does not establish an interference threshold for this Palinoia kit.
When to follow up
Record a non-zero reading beside your other results. Tell your clinician about supplement use if other results seem unexpected or do not fit your symptoms; they can recommend confirmation when appropriate.
On your Palinoia chart
These labels follow the charts supplied with both 14-in-1 kits: the color chart printed on the bottle’s label and the pouch kit’s color chart. Values are in mmol/l.
Outlined colors are inside the chart’s square of ideal results.
Both 14-in-1 charts print 0 inside the square, then 0.6, 1.4, 2.8 and 5.0 mmol/l. Beside the 0 square the chart reads: “To get correct readings on all other tests, you will want this to read zero.”
Compare your pad with your own kit’s chart, in good light, at the time printed beside its row: 40 seconds on both 14-in-1 charts. On-screen colors are illustrative and should not be used to read a sample.
Sources & further reading
- Lee et al.: Vitamin C interference in urine dipstick tests
- Unic et al.: Ascorbic acid and urine chemistry screening
- Palinoia 14-in-1 results and reading-time guide
Chart labels and reading times follow the bottle label and the pouch kit’s color chart. General information is educational; a strip result alone does not diagnose a condition.